A Reliable Quantification Method for Trimethoprim in Pharmaceutical Samples by HILIC-HPLC
DOI:
https://doi.org/10.37506/ijfmt.v15i1.13761Keywords:
Trimethoprim, HILIC, Tablet, syrup, UV-detection; toxicityAbstract
Trimethoprim is sometimes used in conjunction with sulfonamides for drug therapies Herein. A practicable
method has been developed to determine trimethoprim in pharmaceutical formulations by hydrophilic
interaction liquid chromatography coupled with an ultraviolet detector. The development and investigation
of a new HLIC assay of trimethoprim. The separation was made in the HALO-HILIC column using a
NaOAc/HAc buffer (40 mM-pH 4.75) with acetonitrile (10:90) (v/v) as mobile phase and was quantified
by UV detection at 280 nm. The limit of quantification was about 0.0242 ppm and the limit of detection
0.008 ppm. The calibration curves were linear in the investigated range (0.01-3 ppm). The recovery of
trimethoprim was 99-102.5%. The method was applied in pharmaceutical formulations.
Published
How to Cite
Issue
Section

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Re use and mixing of content policy- We follow Creative Commons Licence Policy. We follow CC BY. Please refer below for all details
https://creativecommons.org/licenses/
CC BY
This license lets others distribute, remix, adapt, and build upon our work, even commercially, as long as they credit us for the original creation.
- The journal allows readers to read, download, copy, distribute, print, search, or link to the full texts of its articles and allow readers to use them for any other lawful purpose.
- The journal allows the author(s) to hold the copyright without restrictions.
- The journal allows the author(s) to retain publishing rights without restrictions